Information was received from a foreign healthcare professional (hcp) via a manufacturer representative regarding a procedure involving an intrathecal catheter for an infusion pump system.It was reported the catheter being implanted did not give back csf, so they had to explant it and then could see the catheter was damaged.It was noted as a factor that may have led or contributed to the issue that the surgeon had difficulties to ¿ascend¿ the catheter during the procedure and that there was a question about whether it was damaged by the needle.No diagnostics or troubleshooting was performed.Another catheter was implanted instead.It was indicated that the explanted catheter would be returned for analysis.At the time of the report the issue was resolved, and the patient status was alive without injury.The stated patient¿s medical history was premature and deaf.No further patient complications were reported or anticipated.
|
H3: the returned device was subjected to a series of standard tests that include but is not limited to visual inspection, patency testing, and pressure testing.Analysis identified discoloration on the catheter body which did not affect its performance.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|