PHILIPS ULTRASOUND, INC S8-3T MICRO TEE TRANSDUCER Back to Search Results
Model Number 989605431171
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event Description
A customer reported an s8-3t model transducer was producing poor image quality.There was no injury associated with this event.
 
Manufacturer Narrative
Evaluation of the transducer confirmed the imaging issue as described by the customer.Investigation of the device revealed cracked and chipped beading, scratches on the handle, bite marks on the shaft, damage to the i-tube, sheath, window, tip shells, flex link, and oil separation in the window.The extensive physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S8-3T MICRO TEE TRANSDUCER
Type of Device
S8-3T MICRO TEE TRANSDUCER
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
MDR Report Key10116392
MDR Text Key194313705
Report Number3019216-2020-00032
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00884838073524
UDI-Public(01)00884838073524
Combination Product (y/n)N
PMA/PMN Number
K030455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number989605431171
Device Lot NumberB2HWXX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2020
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-