• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 125 RAD DRILL GDE DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 125 RAD DRILL GDE DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71674002
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/11/2020
Event Type  malfunction  
Event Description
It was reported that during an inspection, the pin of the 125 rad drill guide drop was seen to be missing.No patient was involved.No other complications were reported.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned drop confirms push pin is missing from the device.The missing piece was not returned with the device.The device was manufactured in 2012.The device exhibits signs of significant wear and usage.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
125 RAD DRILL GDE DROP
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10117066
MDR Text Key193942976
Report Number1020279-2020-01977
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010570970
UDI-Public3596010570970
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71674002
Device Lot Number12MM03527
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-