MAUDE Adverse Event Report: NIHON KOHDEN ORANGEMED, INC. NIHON KOHDEN NKV-550 SERIES VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN ORANGEMED, INC. NIHON KOHDEN NKV-550 SERIES VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number NKV-550-U
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problem Respiratory Failure (2484)
Event Type  malfunction  
Manufacturer Narrative
A software bug was identified regarding the numeric monitors during cpap mode.In all cases the patient is receiving the correct respiratory support through cpap as set by the clinician and the displayed waveforms are graphically correct.Because the patient respiratory support is not affected and the graphic monitors function correctly, this software bug did not cause any adverse impact to the patient.
 
Event Description
It was reported that during spontaneous cpap with 9 cmh2o the displayed monitor values were reading very high airway pressures, it wasn't recording tidal volumes, peep, respiratory rate.The patient was stable and oxygenated at all times.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIHON KOHDEN NKV-550 SERIES VENTILATOR SYSTEM
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NIHON KOHDEN ORANGEMED, INC.
1800 e. wilshire ave.
santa ana CA 92705
Manufacturer (Section G)
NIHON KOHEN ORANGEMED, INC.
1800 e. wilshire ave.
santa ana CA 92705
Manufacturer Contact
sheryl higgins
1800 e. wilshire ave.
santa ana, CA 92705
949502644
MDR Report Key10124812
MDR Text Key209400299
Report Number3014631252-2020-00001
Device Sequence Number1
Product Code CBK
Combination Product (y/n)Y
Reporter Country CodeEI
PMA/PMN Number
K181695
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/01/2005,06/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNKV-550-U
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received05/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight81
-
-