NIHON KOHDEN ORANGEMED, INC. NIHON KOHDEN NKV-550 SERIES VENTILATOR SYSTEM; VENTILATOR, CONTINUOUS, FACILITY USE
|
Back to Search Results |
|
Model Number NKV-550-U |
Device Problem
Application Program Problem: Parameter Calculation Error (1449)
|
Patient Problem
Respiratory Failure (2484)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A software bug was identified regarding the numeric monitors during cpap mode.In all cases the patient is receiving the correct respiratory support through cpap as set by the clinician and the displayed waveforms are graphically correct.Because the patient respiratory support is not affected and the graphic monitors function correctly, this software bug did not cause any adverse impact to the patient.
|
|
Event Description
|
It was reported that during spontaneous cpap with 9 cmh2o the displayed monitor values were reading very high airway pressures, it wasn't recording tidal volumes, peep, respiratory rate.The patient was stable and oxygenated at all times.
|
|