Additional information was added to d4 lot #, h3, h4 and h6.D4: lot #: the customer reported a suspect lot gr20c06017.H4: the suspect lot was manufactured between april 3, 2020 and april 5, 2020.H10: only two (2) actual samples were received for evaluation (one activated and one in the inactivated position) attached to vials and solution bags.Visual inspection was performed which observed fluid on the outside of the bags.Functional testing was performed for both samples using a new solution bag (returned bags did not have enough solution) and the original vials.The samples functioned as expected with no leaks observed.No additional defects were observed on samples.The reported condition was not verified for the returned samples during functional testing.The remaining two (2) samples were not received for evaluation; therefore, a device analysis could not be completed for those samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|