MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Blower (738); Complete Blockage (1094); Noise, Audible (3273)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

Date of event: (b)(6) 2020. Date of report: 05june2020.

 
Event Description

It was reported that the unit declared an occlusion alarm during set up. There was no patient involvement. The manufacturer's remote service technician performed troubleshooting with the customer. The customer removed the top of the unit and tapped on the blower assembly. The customer reported that the blower started to run and was very noisy. The technician recommended that the customer replace the blower assembly.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameV60 VENTILATOR
Type of DeviceVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
carlsbad CA
Manufacturer Contact
bill cole
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key10125855
MDR Text Key194779900
Report Number2031642-2020-01987
Device Sequence Number1
Product Code MNT
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation BIOMEDICAL ENGINEER
Type of Report Initial,Followup
Report Date 05/18/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/05/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator OTHER
Device MODEL NumberV60
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received05/18/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured08/24/2010
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-