Model Number 2426-0007 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Although requested, device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
|
|
Event Description
|
It was reported that while troubleshooting a repeated occlusion alarm during an infusion of insulin that had ben infusing for 5 hours in the operating room, it was noticed that the tubing set developed a bulge in the pump segment tubing.The tubing was reinserted into the pump and the pump appeared to function.The case finished without issue.The tubing was collected after the issue and appeared to be normal at the end of the case.It was reported there was no adverse events.
|
|
Event Description
|
It was reported that while troubleshooting a repeated occlusion alarm during an infusion of insulin that had ben infusing for 5 hours in the operating room, it was noticed that the tubing set developed a bulge in the pump segment tubing.The tubing was reinserted into the pump and the pump appeared to function.The case finished without issue.The tubing was collected after the issue and appeared to be normal at the end of the case.It was reported there was no adverse events.
|
|
Manufacturer Narrative
|
Additional info: d.11;g.5.The customer¿s report of a balloon in the tubing set was confirmed during pressure testing.Visual inspection did not observe any anomalies.Testing concluded no occlusion alarms or ballooning/bulge were observed on the pumping segment during an infusion device testing.However, during pressure testing a balloon bulge was observed on the silicone segment below the upper fitment.Further visual inspection under magnification found the walls of the silicone tubing segment to be concentric.The root cause could not be definitively determined.
|
|
Search Alerts/Recalls
|