• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH NON CONNECT 4TH ED DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH NON CONNECT 4TH ED DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66802134
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 05/12/2020
Event Type  malfunction  
Event Description
It was reported that the device did not start-up correctly and was not free from critical errors because the main board was defective and the software was corrupt.No patient harmed and no injury reported because this was found during service and repair.
 
Manufacturer Narrative
H10.H3, h6: the device, intended for use in treatment, has been returned for evaluation.A visual inspection found no defects.The functional evaluation found the device was unable to start up free from critical errors, establishing a relationship with the event reported.The root cause has been identified as a software failure.A review of manufacturing records for the reported lot/batch, found no non-conformances or anomalies during production.The device/product met all specifications upon release into distribution.Complaint history for the reported events has been reviewed, revealing further instances.No further actions are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith + nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH NON CONNECT 4TH ED DEVICE
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10126841
MDR Text Key194327361
Report Number8043484-2020-00666
Device Sequence Number1
Product Code OMP
UDI-Device Identifier05000223494988
UDI-Public5000223494988
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66802134
Device Lot NumberKTCD190291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2020
Date Manufacturer Received02/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-