Model Number 209999
Device Problems Calibration Problem (2890); Inappropriate Tactile Prompt/Feedback (4023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2020
Event Type  malfunction  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical. A supplemental report will be submitted when additional information becomes available.
 
Event Description
After making the distal femoral, posterior and posterior chamfer cut the arm would not engage into the haptics. It acted like it was trying to engage but it would either not move or move into an unrealistic position. The femoral array was not bumped. It would not let me engage in rio setup or rio reg. The robot was shut down and re-homed and re-registered and we were able to continue out bone prep cuts without any issues. Case type: tka. Surgical delay<
=
15 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name3.0 RIO ROBOTIC ARM - MICS
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
daniela davila
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key10130898
MDR Text Key194402590
Report Number3005985723-2020-00212
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number209999
Device Catalogue Number209999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/17/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 06/08/2020 Patient Sequence Number: 1
-
-