Date of event: (b)(6) 2020.Date of report: 08jun2020.
 
Event Description
The customer reported the unit went vent inoperable at turned on.The customer reported that the unit was not in use on patient.The customer contacted product support and requested for service repair.Product support advised customer to replaced power supply and the issue continued, the customer swapped in a central processing unit (cpu) board and the issue continued.The customer swapped in a main harness.The customer found that the unit would start in diagnostics with the screen blank.Tried a video graphics array (vga) card and the issue remained the same.Continued troubleshooting and found that the customer did not have the shroud with the direct current (dc) to dc converter connected.The customer reconnected the dc to dc converter and noted that the unit would now operate correctly.The customer replaced the main cable harness to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESPRIT VENTILATOR
Type of Device
VENTILATOR
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
carlsbad CA
Manufacturer Contact
bill cole
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key10131195
MDR Text Key194793739
Report Number2031642-2020-02004
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K981072
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV1000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/16/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-