This report has been identified as b.Braun medical inc.Internal report number (b)(4).A b.Braun technician completed a technical safety check on the machine per the service manual checklist.A self-test was run and a mock therapy was performed, both with passing results.The machine was disinfected and released for use.The evaluation of the data record of the dialog+ machine shows that the therapy ran for 1 hr 14 minutes without any issue.Then, the alarms "abpm: sys press.Is too low" and "abpm: dia press.Is too low" were triggered because the patient's blood pressure had dropped to 49:11 mmhg (pulse 80 bpm).The operator acknowledged the alarms and activated minimal ultrafiltration.This was obviously the time when the patient developed the shortness of breath.Seven minutes later the blood pressure was 133:49 mmhg (pulse 80 bpm).The operator deactivated the minimal ultrafiltration and activated it again shortly thereafter.The alarm "abpm: dia press.Is too low" was triggered again because the patient's blood pressure was 174:29 mmhg (pulse 93 bpm).Finally, the blood was returned to the patient 1 hr 30 minutes after therapy start.The evaluation of the data record of the dialog+ machine does not show a malfunction.The dialysis machine operated as intended.If additional pertinent information becomes available a follow-up report will be filed.
|