Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Gas/Air Leak (2946)
Patient Problem No Patient Involvement (2645)
Event Date 05/19/2020
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws. The distributor performed a checkout of the equipment and confirmed the reported complaint. The o2 pressure switch was replaced, and the unit was returned to service.
 
Event Description
The hospital reported that the tidal volume was not being achieved due to a leak greater than 4. 5lpm. There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE 7100
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH 214028
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key10143426
MDR Text Key194810588
Report Number9710602-2020-00182
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial
Report Date 06/11/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received05/19/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured05/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-