(b)(4).The actual device was not returned; however, the customer provided one photo for evaluation.The photo displayed two endurance catheters: one contained no damage and one appeared to have a sheared tip.Evidence of use was observed on the body of the damaged catheter.However, a full visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.The instructions-for-use provided with this kit cautions the user, "do not attempt to advance needle back into catheter after catheter is partially threaded off needle to reduce risk of catheter damage." the customer report of a separated catheter tip was confirmed by visual examination of the customer supplied photo.The image shows an endurance catheter with evidence of use and a sheared tip.However, root cause verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|