• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problem Self-Activation or Keying (1557)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/13/2020
Event Type  malfunction  
Event Description
It was reported that during a shoulder scope procedure, the mdu starts and stops on its own.Backup device was available to complete the procedure, no delay and no patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3, h6: the reported device, used in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The housing and the gearbox were removed from the motor.The gearbox was found to be jammed.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A blade stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10151359
MDR Text Key195389038
Report Number1643264-2020-00342
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2020
Date Manufacturer Received07/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-