SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 72200616 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/27/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the motor of the mdu stalled.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Product failed functional testing with blade stall error.Cause of errors is a corroded motor/gearbox.The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures.
|
|