Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 05/21/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 02166, 0001825034 - 2020 - 02167, 0001825034 - 2020 - 02168, 0001825034 - 2020 - 02169, 0001825034 - 2020 - 02171.
 
Event Description
It was reported circulated items were investigated and identified debris in sterile packages.No patients were involved.No further event information is available at the time of this report.
 
Event Description
Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The event reports that debris was identified inside the sterile packaging.Evaluation of the returned product/photographs provided confirmed the following: lot #2920983, 3581704, 2563642, 3684090: debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Lot #2725334 damage to sterile blister and debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Lot #3391428 damage to sterile blister and debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging and the porous coating inside the sterile barrier.The reported event is confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage causing the foam packaging to become abraded and shed.As part of the capa process, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS HO 15X150MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10152279
MDR Text Key195097380
Report Number0001825034-2020-02170
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K101086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model NumberN/A
Device Catalogue Number51-104150
Device Lot Number2563642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-