It was reported by affiliate via email that during an unknown procedure, while using a truespan meniscal repair system peek 12 degree, when the point was to be made with the suture, the implant trigger was pulled once and two points came out at the same time, so implant could not be used correctly.No surgical delay or patient consequence reported.
|
Product complaint # (b)(4).Investigation summary: according to the information provided, it was reported that during an unknown procedure, while using a truespan meniscal repair system peek 12 degree, when the point was to be made with the suture, the implant trigger was pulled once and two points came out at the same time, so implant could not be used correctly.Upon visual inspection of the photo, it was observed only the suture and the truespan gun.The silicone sleeve was broken nearest the needle.Besides, the suture is frayed on the distal.Since the truespan gun was not returned, it is not possible to test its functionality.The photo not provide evidence of the defective into device, therefore the complaint reported cannot confirmed.However, it cannot be determined at what point in time this failure occurred.As a result, we cannot determine a root cause for the reported failure.The possible root cause for the suture damage can be related to rough deployment causing that it had to be cut and remove; thus, it cannot be conclusively affirmed.Hands on analysis should provide more evidence to be able to discern a root cause.A manufacturing record evaluation was performed for the finished device lot number:6l43462, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|