• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE ANALYZER, PACEMAKER GENERATOR FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2020
Event Type  Malfunction  
Manufacturer Narrative

Product analysis: analysis of the mobile app logs indicated the implantable device application had incorrect data. Analysis concluded that the mobile programmer miscalculated the longevity and led to an overestimated value. Further action was not required because an existing corrective action or investigation is applicable. If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the caller was wondering why there was a discrepancy between two types of programmers when the caller interrogated a device and received a difference in the device estimated longevity between both programmers. The device remains in use. There was no patient complications reported with this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC BASE
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10157123
MDR Text Key195230312
Report Number3004593495-2020-00621
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 04/16/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/16/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/19/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured12/07/2018
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number2182208-033021-002-C

-
-