Model Number SGC0301
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis (2100)
Event Date 05/27/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record did not identify any manufacturing nonconformities issued to the reported lot that could have contributed to the reported event.Based on the information reviewed and without the device to analyze, a definite cause for the reported thrombosis could not be determined.The reported patient effect of emboli (thrombus) as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
This is filed to report thrombus.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4+.The steerable guide catheter (sgc) was inserted without issues; however, while at the septum, thrombus was noticed in the right atrium (ra) attached to the dilator.It was noted that the activated clotting time (act) remained above 250 throughout the procedure.The sgc was removed and a 14f sheath was inserted in an attempt to aspirate the thrombus.The thrombus was successfully removed.The sgc was attempted to be re-prepared; however, the sgc came in contact with the iv pole.Therefore, the sgc was not used and was replaced.Two clips were successfully implanted, reducing mr to a grade of 2+.There was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10158256
MDR Text Key195263169
Report Number2024168-2020-05064
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/21/2021
Device Model NumberSGC0301
Device Catalogue NumberSGC0301
Device Lot Number00121U313
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-