The actual device was not available; however, two (2) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing, including under water pressure testing, was performed and no leakage or blockage was observed.The reported condition was not confirmed on the returned companion samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|