BAXTER HEALTHCARE CORPORATION INTERLINK SYSTEM; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C6401
Device Problems Material Deformation (2976); No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The actual device was not available; however, two (2) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing, including under water pressure testing, was performed and no leakage or blockage was observed.The reported condition was not confirmed on the returned companion samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that three (3) interlink system solution sets would not flow during use with sodium chloride.It was further reported that the end of the tubing was kinked.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SYSTEM
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key10167691
MDR Text Key195581206
Report Number1416980-2020-03431
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412002118
UDI-Public(01)00085412002118
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123868
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C6401
Device Lot NumberSR19B18095
Was Device Available for Evaluation? No
Date Manufacturer Received05/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
SODIUM CHLORIDE
-
-