• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS CS2 ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS CS2 ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Crack (1135); Gas Leak (2946)
Patient Problem No Patient Involvement (2645)
Event Date 05/25/2020
Event Type  Malfunction  
Manufacturer Narrative

No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws. A ge healthcare service representative performed a checkout of the system and confirmed the reported issue. The bellows housing was replaced to resolve the reported issue.

 
Event Description

The hospital reported cracked bellows causing a large leak and loss of mechanical ventilation. There was no report of patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAISYS CS2
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key10168706
MDR Text Key197184072
Report Number2112667-2020-01754
Device Sequence Number1
Product Code BSZ
Combination Product (Y/N)N
PMA/PMN NumberK172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,HEALTH PROFESSIONAL,U
Type of Report Initial
Report Date 06/18/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/18/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Was Device Available For Evaluation? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received05/25/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured09/09/2013
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-