MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Electromagnetic Interference (1194); Failure to Interrogate (1332); Communication or Transmission Problem (2896); Insufficient Information (3190)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/23/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information received from the manufacturer representative (rep) reported that the patient underwent a magnetic resonance imaging (mri) without setting the implantable neurostimulator (ins) into mri mode.Since the mri, the ins does not work anymore, interrogation using the communicator or patient programmer is impossible, recharging using antenna-mode without success, and the ins does not give any feedback to the programmer.Factors that may have led or contributed to the issue was the patient having a mri without using mri-mode/switching the ins off.Interventions taken include the replacement of the ins was scheduled.The issue was not resolved and no surgical intervention occurred.Additional information received reported that the patient underwent an magnetic resonance imaging (mri) without using the mri-mode.Leads were implanted subcutaneously.Since the mri, the implantable neurostimulator (ins) does not work anymore.Interrogation using the programmer was impossible and recharging using antenna-mode without success.The ins does not give any feedback to the programmer.Actions taken was a recharge using antenna mode where there was a 98.The issue resolved.
|
|
Event Description
|
Additional information received reported that the ipg recharging was solved.The ipg is working properly again.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|