MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065752914
Device Problem Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/27/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported oscillation failure of the ultrasound handpiece and that the handpiece generated heat during a surgical procedure.The handpiece was exchanged to complete the surgery with no patient harm.
 
Manufacturer Narrative
No further information was able to be obtained from this customer.However, a handpiece was returned for evaluation.The manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.A visual assessment of the returned sample found no visual non-conformities.The handpiece was connected to a resistance test box, where it¿s input and output impedance were found to be within specification.The returned sample was connected to a calibrated system.The handpiece tuned successfully and completed a five minute burn-in test with the system set at 100% ultrasonic and torsional power.During the burn-in test the temperature of the hp was measured and was found to be within specifications.The handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the handpiece to meet specifications.The handpiece was found to meet specifications.Therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10171253
MDR Text Key196395967
Report Number2028159-2020-00449
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752914
Device Lot Number13964A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2020
Date Manufacturer Received07/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM
-
-