MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Failure to Interrogate (1332); Overheating of Device (1437); Charging Problem (2892); Communication or Transmission Problem (2896); Connection Problem (2900); Insufficient Information (3190)
|
Patient Problem
Burning Sensation (2146)
|
Event Date 03/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger; product id: 97755, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot#: (b)(4), udi#: (b)(4).Analysis of the recharger (serial# (b)(4)) found the recharger was scrapped due to insulation pulling away at the donut end.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins).It was reported there the patient started having problems with the system in early (b)(6) 2020.The patient stated they had experienced a variety of unknown errors.The patient was directed to reset their controller.The patient clarified the performance of the equipment seemed to be deteriorating but the main issue was the recharger.It was noted there was a cord/cable issue.The recharger would get really warm and wouldn't charge the implant.The recharger couldn't find the implanted device and it over heated, was super-heated and was burning them.The patient stated they would go through passive recharge mode which sent power to the paddle and that was what was burning him.No further symptoms or complications were reported or anticipated.
|
|
Manufacturer Narrative
|
Product id: 97755, serial# (b)(4), product type: recharger; product id: 97755, serial# (b)(4), product type: recharger.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
G4: review of this mdr and/or additional information received shows that the previous reported initial aware date of (b)(6) is no longer applicable.The corrected initial aware date is (b)(6), and the original report was on time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|