DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problems Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Patient Problems Pain (1994); No Code Available (3191)
Event Date 06/09/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4). If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that patient presented with a painful hip after hearing a popping sound. Surgeon believed the poly liner may have disassociated and elected to revise the hip. The poly liner was disassociated and many of the locking tabs were worn away. Surgeon diligently inspected the locking mechanism for damaged and elected to place a new poly in the existing cup and exchange the 28 b ceramic head for a ts ceramic 36 head. Doi: (b)(6) 2012. Dor: (b)(6) 2020; affected side: right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN HIP ACETABULAR CUP
Type of DeviceHIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key10184465
MDR Text Key196133770
Report Number1818910-2020-14277
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/23/2020 Patient Sequence Number: 1
Treatment
ALTRX NEUT 28IDX52OD; DELTA CER HEAD 12/14 28MM +5
-
-