RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 23jun2020.
|
|
Event Description
|
It was reported there is was an error, however the exact nature of the error is unknown.The issue happened while the device was in use, however there was no reported harm to the patient or user.
|
|
Manufacturer Narrative
|
G4: 06nov2020; b4: 09nov2020.It was reported there was a proximal sensor auto zero failure.The device was on a patient at the time of the issue and the ventilator was swapped out.No harm was reported.A circuit test was done and the result was port failure.The customer replaced the flow sensor assembly and the issue was resolved.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|