|
Model Number 9388 |
Device Problem
Positioning Problem (3009)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/27/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device is a combination product device evaluated by mfr: promus element plus,mr,ous 3.50x24mm stent delivery system was returned for analysis.A visual examination of the stent found stent damage.Distal stent struts were lifted from their crimped position.The undamaged stent outer diameter was measured using a snap gauge and the result was within maximum crimped stent profile measurement.The balloon was reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the shaft polymer extrusion found no issues.An examination (visual and via scope) found tip damage.No other issues were identified during the product analysis.
|
|
Event Description
|
Reportable based on device analysis completed on 28may2020.It was reported that stent difficult to position occurred.The more than 90% stenosed target lesion was located in the left anterior descending artery.After the lesion was crossed with a guidewire and pre-dilated with a balloon, a 3.50x24mm promus element plus drug-eluting stent was advanced for treatment.However, during procedure, the stent could not positioned accurately in the lesion due to severe tortuosity in the vessel.The device was removed and the procedure was completed with a different device.There were no patient complications reported and the patient was stable.However, returned device analysis revealed a stent damage.
|
|
Search Alerts/Recalls
|
|
|