Model Number N/A
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Date 06/04/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Pma/510k: k081047; k123188; k133786.Evaluation and investigation is in process.Once the investigation is complete, a supplemental / final medwatch will be filed.Evaluated by external contractor.
 
Event Description
It has been reported that the unit was showing full when both cylinders were empty.The event timing was during cleaning.There was no harm to the patient.No adverse events were reported as a result of this malfunction.
 
Event Description
There is no additional information.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Current repair.Product evaluation: product review of the ultra cart serial number (b)(6) by zimmer biomet certified service repair technician on 08 june 2020 revealed that there was no problem found, technician could not duplicate level sensor issue.Product repair: repair of the device was not performed because there was no problem found.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX DUO HIGH FLUID CART - REFURBISHED
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
MDR Report Key10194568
MDR Text Key197559526
Report Number0001954182-2020-00026
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
PMA/PMN Number
K133786
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULDU500R
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A