DORNOCH ULTRAFLEX DUO HIGH FLUID CART - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Pma/510k: k081047; k123188; k133786.Evaluation and investigation is in process.Once the investigation is complete, a supplemental / final medwatch will be filed.Evaluated by external contractor.
|
|
Event Description
|
It has been reported that the unit was showing full when both cylinders were empty.The event timing was during cleaning.There was no harm to the patient.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Current repair.Product evaluation: product review of the ultra cart serial number (b)(6) by zimmer biomet certified service repair technician on 08 june 2020 revealed that there was no problem found, technician could not duplicate level sensor issue.Product repair: repair of the device was not performed because there was no problem found.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event cannot be confirmed.
|
|