NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number NX1000-1
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Death (1802)
Event Date 05/28/2020
Event Type  Death  
Manufacturer Narrative
The device was received for evaluation and successfully passed the prime and alarms test confirming proper functionality of the device and sensors.Available log files were retrieved and analyzed which showed device performance was without incident and was unremarkable.A review of the device history record confirmed the product was released meeting all product design specifications and quality criteria.All information supports the device was functioning as designed and intended with no malfunction.Udi: (b)(4).
 
Event Description
A report was received on (b)(6) 2020 from the home therapy nurse (htn) of a (b)(6) female patient approved for performing solo home hemodialysis treatments, with a medical history of end stage renal disease and diabetes, stating the patient expired while performing their hemodialysis treatment on (b)(6) 2020.Treatment details and cause of death were requested and have not been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key10197222
MDR Text Key196545045
Report Number3003464075-2020-00034
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K050525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/01/2005,06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2020
Is this an Adverse Event Report? Yes
Device Operator Lay User/Patient
Device Model NumberNX1000-1
Device Catalogue NumberCYC-D2E W/ SERVICE COMPUTER
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received05/29/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age71 YR
Patient Weight73
-
-