Model Number CD3357-40C
Device Problems Difficult to Interrogate (1331); Inappropriate or Unexpected Reset (2959)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/09/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented in clinic for a follow up.Upon interrogation, the device was found in backup mode, caused by the home monitor and the device attempting to communicate.The device was restored.When continuing the interrogation, the device began to have trouble communicating using inductive telemetry; however, using the rf antenna resolved the issue and all values were found to be normal.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10197760
MDR Text Key196564457
Report Number2017865-2020-08235
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public05414734508186
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberA000037627
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
-
-