Model Number 479878
Device Problems High impedance (1291); Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/08/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a pacing failure was observed on the his lead.The lead was explanted and replaced with an right ventricular (rv) lead.During the lead replacement procedure, the left ventricular (lv) lead exhibited a high, undefined pacing impedance when reconnected to the cardiac resynchronization therapy pacemaker (crt-p).It was suspected that there was a loose connection.An x-ray was taken, blood that had adhered to the lead was wiped out, and the crt-p was explanted and replaced.When the lv lead was connected to the new device, all measurements were normal.The lv lead remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STABILITY QUAD MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10199834
MDR Text Key196573740
Report Number2649622-2020-12195
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/13/2022
Device Model Number479878
Device Catalogue Number479878
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2020
Date Device Manufactured04/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
W4TR04 CRTP, 5076-45 LEAD, 383069 LEAD
Patient Age77 YR
-
-