• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 26jun2020.
 
Event Description
It was reported that the unit failed alarm testing and that the test cable resistance was too high.There was no patient involvement.The manufacturer's remote service technician performed troubleshooting with the customer.The technician recommended that the customer use a steel wool pad to dull the chrome finish on the cable to see if this resolves the issue.The customer reported that they dulled the test cable and the issue was resolved.The unit was returned to service.
 
Manufacturer Narrative
G4: 15jul2020 b4: (b)(6) 2020 communication attempt made on: (b)(6)2020 @ 6:05am email from (b)(6.The customer stated the issue with the test cable was resolve the device was back in service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10202735
MDR Text Key197109872
Report Number2031642-2020-02198
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 06/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-