RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

Date of event: (b)(6) 2020. Date of report: 26jun2020.

 
Event Description

It was reported that the unit failed alarm testing and that the test cable resistance was too high. There was no patient involvement. The manufacturer's remote service technician performed troubleshooting with the customer. The technician recommended that the customer use a steel wool pad to dull the chrome finish on the cable to see if this resolves the issue. The customer reported that they dulled the test cable and the issue was resolved. The unit was returned to service.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameV60 VENTILATOR
Type of DeviceVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key10202735
MDR Text Key197109872
Report Number2031642-2020-02198
Device Sequence Number1
Product Code MNT
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation BIOMEDICAL ENGINEER
Type of Report Initial,Followup
Report Date 06/09/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/26/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator NO INFORMATION
Device MODEL NumberV60
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/09/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured12/19/2012
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-