• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL INFLOW TUBING FMS VUE 24PK; SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL INFLOW TUBING FMS VUE 24PK; SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE Back to Search Results
Model Number 284508
Device Problem Fluid/Blood Leak (1250)
Patient Problems No Consequences Or Impact To Patient (2199); Not Applicable (3189)
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the affiliate in (b)(6) that a water leak occurred during the installation of tubing.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.There were no reports of any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported by affiliate via email that a water leak occurred during the installation of tubing (inflow tubing fms vue 24pk).The complaint device was received and evaluated.Visual inspection reveals that the tubing was received cutted in two sections, was not complete.In addition, the tubing was found cracked close to the clamp set.The functional test was not performed due to damages in the tubing.The complaint can be confirmed.As mentioned in the note a-4213762, the root cause for the reported failure was due to there is a tear at the level of the injection roller when the device is put in place and for the tubing cutted can be attributed to user cut the device.A manufacturing record evaluation was performed for the finished device [5064] number, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional information: b5: subsequent follow-up with the customer, additional information was received.It was reported that the wound of the serum tubing at the injection roller occurred during the assembly of the material before incision.It was reported that there was a tear at the level of the injection roller when the device was put in place.It was reported that the person who installed the tubing did not use it often and has performed two attempts to put it in place because of an interruption but seemed to be a handling problem in this case.Please provide the lot number of the devic.Lot 5064.What was the procedure in which the device was involved? knee ligamentoplasty.When does the event happened? pre-operatively.Was surgery delayed due to the reported event? yes.Please report the time (minutes/hours).Unknown but there was minimal lengthening of procedure time.Was there complication to the patient? no.How was the procedure completed? tubing replacement.Was an alternative product readily available? yes.Udi: (b)(4).The expiration date was unavailable.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d4, h4: the lot number, expiration date and device manufacture date have been updated accordingly.Therefore, udi: (b)(4).H6: method codes: a manufacturing record evaluation was performed and results were obtained as follows: product : 284508 lot number :(b)(4) anomalies or discrepancies (non-conformance) : none.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary according to the information provided, it was reported by affiliate via email that a water leak occurred during the installation of tubing (inflow tubing fms vue 24pk).The complaint device is not being returned, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.Since the complaint device not being returned, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device [5064] number, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFLOW TUBING FMS VUE 24PK
Type of Device
SURGICAL IRRIGATION TUBING SET, GENERAL-PURPOSE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
MDR Report Key10206871
MDR Text Key196846337
Report Number1221934-2020-01680
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705023066
UDI-Public10886705023066
Combination Product (y/n)N
PMA/PMN Number
K951843
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/30/2020
Device Model Number284508
Device Catalogue Number284508
Device Lot Number5064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2020
Date Manufacturer Received09/08/2020
Patient Sequence Number1
-
-