• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 13X146MM 6MM T1 PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 13X146MM 6MM T1 PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 06/09/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Report source: (b)(6). Customer has indicated that the product is in process of being returned to zimmer biomet for investigation. Once the investigation has been completed, a follow-up mdr will be submitted. Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-02558, 0001825034-2020-02559, 0001825034-2020-02560, 0001825034-2020-02561, 0001825034-2020-02562, 0001825034-2020-02563, 0001825034-2020-02565.
 
Event Description
Customer has indicated that the product is in process of being returned to zimmer biomet for investigation. Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4). Reported event was confirmed by examination of the devices/photographs. Device history record (dhr) was reviewed and no discrepancies related to the event were found. The root cause of the reported event is likely to be due to transit damage. Evaluation of the returned product/photographs provided confirmed the following: lot #2901608; 3791225; 3090879; 2955965; 3473486; 3458643: debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier. Lot #3383473; 3560787: debris inside the sterile packaging which is consistent with the appearance of the porous coating and foam debris from the foam packaging inside the sterile barrier. If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly. Zimmer biomet will continue to monitor for trends. Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications. Event is no longer considered reportable, and initial report should be voided.
 
Event Description
Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications. Event is no longer considered reportable, and initial report should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTPRLC 133 T1 PPS HO 13X146MM 6MM T1
Type of DevicePROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10208902
MDR Text Key201157785
Report Number0001825034-2020-02564
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K101086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/28/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number51-104130
Device Lot Number3458643
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2020
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A

-
-