Model Number PM3210 |
Device Problems
Failure to Interrogate (1332); Wireless Communication Problem (3283)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
|
Event Date 06/17/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that patient¿s device was not able to be interrogated.The interrogation was attempted using inductive telemetry, different programmer, clearing the environment to remove effect of electromagnetic interference (emi), magnet, radio frequency telemetry test.All troubleshooting failed.The decision was made to interrogate the device in a different room.If interrogation fails, the device replacement was discussed.Further information was requested but was not received.
|
|
Event Description
|
New information received notes that the device was explanted and replaced.Elective replacement indicator (eri) was also noted.
|
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Manufacturer Narrative
|
The reported field event of failure to interrogate was not confirmed in the laboratory.The device was tested on the bench, and no anomalies were found.
|
|