Model Number N/A |
Device Problems
Loss of or Failure to Bond (1068); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problem
Pain (1994)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it's current location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the patient was revised due to the ulnar component being completely loose.Ulnar bony wear with impending fracture.Resection arthroplasty.All cultures and sonication negative for infection.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device history record (dhr) review was unable to be performed as the part and lot number of the device involved in the event is unknown.Radiographs were provided and reviewed by a health care professional.No action was taken at this time as images do not appear to contain product to be assessed.Patient was also noted to have a nickel allergy, however, documentation was not provided to confirm.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|