MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1070-040
Device Problems Material Rupture (1546); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/05/2020
Event Type  Malfunction  
Manufacturer Narrative

The device was not returned for analysis. A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint. Additionally, a review of the complaint history of the reported lot revealed no other complaints reported from this lot. Based on the information provided, the resistance during advancement and balloon rupture appear to be due to case circumstances. It is likely that the resistance during advancement and rupture occurred due to interaction with lesion calcification. There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.

 
Event Description

It was reported that the procedure was performed to treat a lesion in the heavily calcified, moderately tortuous superficial femoral artery. A 7x40mm armada 35 percutaneous transluminal angioplaty (pta) catheter felt resistance with the anatomy during advancement before the balloon ruptured during the first inflation at 12 atmospheres. An unspecified armada 35 pta was used to successfully complete the procedure. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 35 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10211658
MDR Text Key201563096
Report Number2024168-2020-05421
Device Sequence Number1
Product Code LIT
Combination Product (Y/N)N
Reporter Country CodeGM
PMA/PMN NumberK111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial
Report Date 06/30/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/30/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberB1070-040
Device Catalogue NumberB1070-040
Device LOT Number00224G1
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/09/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured02/24/2020
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-