The company service representative examined the system and did not indicate replicating the reported event.The illuminator module was replaced.The system was then tested and met all product specifications.The illuminator module was received and tested.The evaluation found a nonconforming fiber cable assembly in the illuminator module.The system was manufactured on march 13, 2009.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the nonconforming illuminator module fiber cable assembly.The manufacturer internal reference number is: (b)(4).
|