Model Number CV-190 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the service department of olympus (b)(4), it was found that the endoscopic image of the subject device was not displayed.There was no report of patient injury associated with this event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus europa se & co.Kg (oekg) there was the possibility that the reported phenomenon was attributed to the temporary unstable contact of the video connector socket or temporary failure of the electrical circuit board.If additional information becomes available, this report will be supplemented.
|
|