Model Number MMT-1711K
Device Problem Power Problem (3010)
Patient Problem Hyperglycemia (1905)
Event Date 06/19/2020
Event Type  Malfunction  
Manufacturer Narrative

(b)(4). Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. No conclusion can be drawn at this time. We therefore consider this report complete to the best of our knowledge. The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states. However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.

 
Event Description

Information received by medtronic indicated that the insulin pump was off and there was no power. Customer stated that connection was lost between transmitter and the insulin pump after insulin pump was out of battery for few hours. No harm requiring medical intervention was reported. The insulin pump will not be returned for analysis.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name640G INSULIN PUMP MMT-1711K
Type of DevicePUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
00777-3869
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
  00777-3869
8185464805
MDR Report Key10216160
MDR Text Key197635012
Report Number2032227-2020-139184
Device Sequence Number1
Product Code OYC
Combination Product (Y/N)N
Reporter Country CodeQA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type DISTRIBUTOR,FOREIGN
Reporter Occupation
Type of Report Initial
Report Date 06/30/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received06/30/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device MODEL NumberMMT-1711K
Device Catalogue NumberMMT-1711K
Device LOT NumberHG36DST
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/19/2020
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured02/15/2019
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-