It was reported that there was a hole in the iv tubing.Received from the customer is one used primary set model 2420-0007, lot 20033183.The set configuration was visually inspected as received for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection found a cut in the primary set tubing (p/n 660132), approximately 1 inch below the distal smartsite port.No other anomalies were observed throughout the set configuration.Functional testing was performed by filling a lab iv bag with blue dye water and attaching it to the set allowing fluid to flow through the whole set by gravity.The set leaked from previously observed cut in the tubing.No other leaks were observed throughout the set.No further functional testing could be performed due to the leak from the tubing.A device history record for model 2420-0007, with lot number 20033183, was performed.The search showed that a total of (b)(4), units in 1 lot number were built on 10mar2020.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.The customer¿s report of a hole in the tubing was confirmed.The root cause of the cut damage could not be determined.
|