PLEXUS MANUFACTURING SDN. BHD DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL
|
Back to Search Results |
|
Model Number 5392 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: the analysis was able to confirm the customer comment of the external pulse generators (epg), battery removal test failed the epg turned itself off.The super caps were not soldered properly to main printed circuit board.It was also determined at analysis that the hanger assembly was broken, and one case screw was contaminated.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported the external pulse generators (epg), battery removal test failed, the battery tray was opened, and the operation dropped immediately, regardless of the fully charged battery.The epg turned itself off and needed to be power cycled to return to operation.The epg was returned to service and repair.There was no patient involvement.
|
|