Model Number RT380 |
Device Problems
Fluid/Blood Leak (1250); Leak/Splash (1354)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Ps342105.The complaint rt380 adult evaqua2 breathing circuit is currently en route to fisher & paykel healthcare (f&p) in (b)(6) for evaluation.We will provide a follow-up report upon completion of investigation.
|
|
Event Description
|
A distributor on behalf of a healthcare facility in (b)(6) reported, via a fisher & paykel healthcare (f&p) field representative, that a rt380 adult dual-heated evaqua2 breathing circuit was found to be leaking water.There was no reported patient consequences.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint rt380 evaqua 2 adult breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected.Results: visual inspection of the returned breathing circuit revealed that the tubing has a hole was found about 10cm from the patient end connector.Conclusion: the damage observed was most likely due to puncture by a blunt object.All rt380 adult evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject infant breathing circuits would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit state the following: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
|
|
Event Description
|
A distributor on behalf of a healthcare facility in japan reported, via a fisher & paykel healthcare (f&p) field representative, that a rt380 adult dual-heated evaqua2 breathing circuit was found to be leaking water.There was no reported patient consequences.
|
|
Search Alerts/Recalls
|