• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DENALI FEMORAL SYSTEM; VENA CAVA FILTER Back to Search Results
Catalog Number DL900F
Device Problems Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/12/2019
Event Type  Death  
Manufacturer Narrative
Manufacturing review: the device history record review could not be performed as the lot number is unknown.Investigation summary: the device was not returned for evaluation.Medical records were not provided.The investigation is inconclusive as no objective evidence has been provided to confirm any alleged deficiency with the filter.Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
 
Event Description
It was reported through the litigation process that some time post vena cava filter deployment the patient expired.There was no specific device malfunction reported that may or may not have caused or contributed to the patient¿s death.The cause of the patient¿s death was not provided.No other pertinent patient, device or medical information was provided leading up to or surrounding the event.
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient with multiple trauma.Approximately three years later post filter deployment, it was alleged that the filter perforated beyond the inferior vena cava wall.The device has not been removed and there were no reported attempts made to remove the filter.The patient reportedly expired.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately, three years of post-deployment, a computed tomography of abdomen was revealed infra renal inferior vena cava filter was in place with superior extent of the filter at the level of the left renal vein.Filter tilt measures approximately six degrees.Filter legs extend up to 3 mm beyond the inferior vena cava wall.Therefore, the investigation is confirmed for the perforation of the inferior vena cava (ivc).Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: g2, g3, d4(expiry date: 04/2019), h6(method, conclusion) h11: b5,d1,d4,h6(device, result) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI FEMORAL SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10231536
MDR Text Key197537459
Report Number2020394-2020-04225
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDL900F
Device Lot NumberGFAR0090
Was Device Available for Evaluation? No
Date Manufacturer Received08/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age38 YR
-
-