Product analysis: analysis of the device was able to confirm the customer comment of a low current on the atrial output, the main printed circuit board (pcb) was found to be out of specification.Analysis also noted that the device had one contaminated case screw.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
Additional analysis was carried out during which the device failed atrial output.It was identified that q79 and q80 malfunctioned.Q79 and q80 are each dual bjt matched pairs that are used for a current mirror.The mirror is used to isolate the circuitry used to create the pulse output shape from the patient and provide a high output impedance.A malfunction of the current mirror will result in no pace output for the corresponding channel.Replaced q79 and q80 with known good transistors.Rerun the device through automated test console ¿ passed all tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|