• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 4008
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2020
Event Type  malfunction  
Event Description
It was reported that before treatment the dressings were creased.S+n backup dressings were available, no patient harm reported.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issues highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaint history review was carried out using the lot and part numbers provided, there have been further complaints reported with this failure mode in the past three years.It was reported that the dressings were used for treatment and creasing was found on the film.The returned dressings were disassembled so it was not possible to perform a product evaluation.There are checks in place to prevent creasing of the film.If creasing is identified during in-process checks, it is tabbed and recorded in the fault log.Minor creasing can sometimes occur in the film during the adhesive application process at the start or end of the roll.Again visual checks are in place to identify this.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.However, we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 10X12CM CTN 50
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10240744
MDR Text Key197816122
Report Number8043484-2020-01730
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial,Followup
Report Date 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4008
Device Lot Number1910
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-