• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 10X12CM CTN 50; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 59410882
Device Problem Product Quality Problem (1506)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2020
Event Type  malfunction  
Event Description
When the customer removed the carrier #2 of the iv3000, the film dressing adhered to the carrier #2 and got peeled off from the frame, so it was impossible to use.A backup was available.No delay was reported.The samples including unopened products from the same carton will be returned for investigation.
 
Manufacturer Narrative
We have now concluded our investigation into this complaint.The batch records were reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of its finished product specification.There was no issues recorded in the batch log of the manufacturing records during the manufacturing process that could have caused or contributed to the issue.A complaint history review was carried out using the lot and part numbers provided, there have been further complaints reported with this failure mode in the past three years.The device was intended to be used for treatment, however, issues were experienced during application.The returned samples were functionally evaluated.The reported issue was not identified on any of the twenty evaluated dressings.The reported issue may be caused by delamination.Delamination is a low level intermittent fault which if seen is tabbed for removal during the conversion process.During the manufacturing of the dressing, in process checks are undertaken for this product.All material is manufactured according to the specification and only product meeting the release criteria will be passed on for further processing.We have not been able to confirm a relationship between the event and the device.The investigation has been closed as no problem found.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 WINDOW FRAME 10X12CM CTN 50
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10241040
MDR Text Key197942500
Report Number8043484-2020-01736
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number59410882
Device Lot Number1848
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-