MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
High impedance (1291); Low impedance (2285); Delayed Charge Time (2586)
|
Patient Problem
Therapeutic Effects, Unexpected (2099)
|
Event Date 07/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer via a manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that there was charging inefficiency/above average charging times, impedance/connectivity issues, and inadequate therapy results.Upon interrogating the battery on 7/4, it was discovered that an impedance issue for electrode 9 was >40k.Electrodes 3, 8, and 14 showed avoid.On 7/6 additional impedances showed as possible shorted electrodes.The therapy was programmed to bypass the electrodes of concern.No further complications were reported.
|
|
Manufacturer Narrative
|
Product id: 977a260, serial# (b)(4), implanted: (b)(6) 2020, product type: lead; product id: 977a260, serial# (b)(4), implanted: (b)(6) 2020, product type: lead.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the rep.It was reported that the programming around solved charging issues and was not getting pain relief.Cause was unknown.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the rep and it was reported that it was taking 2 hours to charge the ins.No further complications were reported/anticipated.
|
|