• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U125
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Patient Involvement (2645)
Event Date 02/29/2020
Event Type  malfunction  
Manufacturer Narrative
At this time, the product has not been returned.
 
Event Description
It was reported that this device recorded a code 1003 during pre-implant interrogation, presumed to be exposure to cold weather.In a different situation would be likely to compromise critical therapy leading to death/serious injury.No patient involvement.
 
Event Description
It was reported that this device recorded a code 1003 during pre-implant interrogation, presumed to be exposure to cold weather.There was no patient involvement at the time.The product was returned for analysis.
 
Manufacturer Narrative
This supplemental report was filled to update h10 field with product analysis results.In consequence, h6 fields were also updated after the product analysis.B5 field was updated.Upon receipt at our post market quality assurance laboratory, an evaluation of the device was performed.Review of the device memory confirmed that a low voltage alert, code 1003, was recorded and that the device recorded low temperature readings prior to setting the low voltage alert.If this model device is stored in environments where temperatures can drop below the recommended storage temperature, battery voltage readings that are low enough to set a low voltage alert may result.This appears to be the case with this device.The battery did recover after the device was removed from the cold.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key10253624
MDR Text Key198381841
Report Number2124215-2020-11486
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2020
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number709236
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2020
Date Manufacturer Received07/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-