BOSTON SCIENTIFIC CORPORATION VALITUDE CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U125 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/29/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the product has not been returned.
|
|
Event Description
|
It was reported that this device recorded a code 1003 during pre-implant interrogation, presumed to be exposure to cold weather.In a different situation would be likely to compromise critical therapy leading to death/serious injury.No patient involvement.
|
|
Event Description
|
It was reported that this device recorded a code 1003 during pre-implant interrogation, presumed to be exposure to cold weather.There was no patient involvement at the time.The product was returned for analysis.
|
|
Manufacturer Narrative
|
This supplemental report was filled to update h10 field with product analysis results.In consequence, h6 fields were also updated after the product analysis.B5 field was updated.Upon receipt at our post market quality assurance laboratory, an evaluation of the device was performed.Review of the device memory confirmed that a low voltage alert, code 1003, was recorded and that the device recorded low temperature readings prior to setting the low voltage alert.If this model device is stored in environments where temperatures can drop below the recommended storage temperature, battery voltage readings that are low enough to set a low voltage alert may result.This appears to be the case with this device.The battery did recover after the device was removed from the cold.
|
|