Per the instructions for use (ifu), valve malposition requiring intervention is a known potential adverse event associated with the transcatheter valve replacement (tvr) procedure.There are multiple patient and procedural factors that alone or in combination can cause or contribute to malposition, including improper positioning prior to deployment, poor image intensifier angle, poor coaxial alignment of the valve and delivery system, severe septal hypertrophy, minimally or bulky/severely calcified leaflets, preserved ejection fraction, significant mitral annular calcification (mac), loss of pacing capture, rapid deployment, release of stored tension during deployment, and movement of the delivery system by the operator.Deployment of the sapien 3 valve too atrial has the potential to contribute to suboptimal coaptation of the sapien 3 valve leaflets and cause valve insufficiency; and lead to embolization of the prosthesis.The thv training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Physicians are extensively trained by edwards before they are qualified to use the sapien 3 thv.Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals and proctored procedures.The correct alignment and positioning of the device at the point of deployment is emphasized as a key factor to the placement and fixation of the device.Operators are also instructed to use fluoroscopy as the primary method of visualization for positioning and deployment.In patients with high-risk anatomical features for aortic malposition (i.E.Minimal leaflet calcification, severe septal hypertrophy), bav may provide indication of potential balloon movement during valve deployment.Per the instructions for use (ifu), paravalvular leak (pvl) is a potential adverse event associated with bioprosthetic heart valves and the transcatheter valve replacement (tvr) procedure.The patient screening manual and the procedure didactic identify several procedural and anatomical factors which could contribute to pvl, including device malposition, inaccurate measurement of the native valve annulus, uneven distribution of calcium on the native valve, bulky or severe calcification, an elliptical annulus shape and valve under-sizing.Physicians are extensively trained by edwards before they are qualified to use the sapien 3 thv.The patient screening manual instructs the operator on proper aortic valve and root assessment, including the use of echo, aortogram and ct to appropriately measure the annulus diameter, content and distribution of calcium, and leaflet characteristics.Contraindications, important considerations when assessing the valve, and choosing the proper thv are also discussed.The thv training manuals also instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.Device preparation, approach, deployment, imaging, procedure-specific training manuals and proctored procedures are also included.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.Investigation results suggest/indicate in addition to the procedural factors (manipulation of the devices), patient factors not provided may have contributed to the too atrial position of the initial valve, pvl and subsequent deployment of a second valve.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
|
As reported by our affiliate in (b)(6), during a mitral valve replacement (tmvi) procedure, a 29mm sapien 3 valve was deployed in the mitral position by the transfemoral approach.During the procedure, valve was implanted too atrial.The valve was functioning well post deployment; however, paravalvular leak (pvl) was observed.A second 29mm sapien 3 valve was successfully implanted, resolving the pvl.The patient was stable after the procedure.
|